You are currently viewing Crane EE-9002 Double Electric Cordless Rechargeable Breast Pump User Manual

Crane EE-9002 Double Electric Cordless Rechargeable Breast Pump User Manual

Crane logo

Crane EE-9002 Double Electric Cordless Rechargeable Breast Pump User Manual

Crane-EE-9002-Double-Electric-Cordless-Rechargeable-Breast-Pump-User-Manual-Image

IMPORTANT SAFETY INSTRUCTIONS

DANGER: To reduce the risk of electrocution

  •  ALWAYS unplug electrical product immediately after use.
  •  DO NOT USE while bathing.
  •  DO NOT place or store this product where it can fall into a tub or sink.
  • DO NOT place in or drop into water or other liquid.
  • DO NOT reach for product if it has fallen into the water. Unplug immediately
  •  NEVER immerse pump unit or AC adaptor in water or attempt to sanitize in a microwave

WARNING: To reduce the risk of burns, electrocution, fire or injury to persons:

  •  Use this product only for its intended use as described in these instructions.
  •  Breast pumps are personal care items for one user only.
  • Danger of electrocution! Keep the device dry, never immerse in water or other liquids. Do not run water over the pump
  • Keep the power adaptor away from heated surfaces.
  • Do not reach for product if it has fallen into water. Unplug immediately
  • Do not use outdoors with AC adaptor, or operate where aerosol spray products are being used, or oxygen is being administered.
  •  Close supervision is necessary when being used near children

Functions & Features

  1. Double sided pumping: Your pump includes everything you need for double pumping. Pumping from both sides at once saves you time and supports better milk supply.
  2.  A fully closed system with anti-back flow.
  3.  Ergonomic milk collection kit: Silicone 3D Cushions allow for a more comfortable fit and help to apply gentle pressure, better suction and massage action to stimulate lactation while pumping.
  4.  2-Phase expression- Stimulation and Expression
  5.  9 Adjustable suction levels
  6.  Control in your hands: With a flexible range of modes, levels and dual pumping options, the Crane double electric breast pump lets you find the ideal pumping combination.
  7.  Touch control: Easy use and & wipe clean controls
  8.  30 min count-up timer: Time your current pumping session up to half an hour.
  9.  All parts are BPA Free.
  10.  Portable, compact design: Gives you and baby more flexibility to pump whether at home or on the go. Your Crane double electric breast pump comes complete with a built in rechargeable li-ion polymer battery.

Parts

Other available parts
The Silicone 3D cushions fit around the nipple and come in a range of sizes, you’ll find the most commonly used size in your Crane breast pump box. Your nipple will expand while milk is being expressed. If you find the standard sized Silicone 3D cushion is too small or too large, other sizes are available from your Crane retailer. See the back of the instructions for information on how to measure your nipple for the correct sized cushion.

Crane-EE-9002-Double-Electric-Cordless-Rechargeable-Breast-Pump-User-Manual-Image-1

Crane-EE-9002-Double-Electric-Cordless-Rechargeable-Breast-Pump-User-Manual-Image-2

Assembling Your Crane Pump

MILK Collection Kit
Insert the silicone 3D cushion into the breast shield body. Ensure the edge of the 3D cushion Is attached to the breast shield body.

Crane-EE-9002-Double-Electric-Cordless-Rechargeable-Breast-Pump-User-Manual-Image-3

Secure valve
Once the silicone 3D cushion is inserted Into the breast shield body, gently push Upwards on the valve to make sure it is Securely attached to the inside body of The breast shield.

Crane-EE-9002-Double-Electric-Cordless-Rechargeable-Breast-Pump-User-Manual-Image-4

Diaphragm and Cap
Carefully insert the diaphragm into the top of the breast shield body. Make sure the edge is flush and the seal is good.
Screw the cap on tightly to prevent Any air from leaking.

Crane-EE-9002-Double-Electric-Cordless-Rechargeable-Breast-Pump-User-Manual-Image-5

Assemble the Bottle
Place the assembled breast shield body onto the bottle. Screw the bottle onto the breast shield body.

Crane-EE-9002-Double-Electric-Cordless-Rechargeable-Breast-Pump-User-Manual-Image-6

Connecting the Tubing Connector Set

Connect one end of the Main Tubing to the single opening of the Y- Connector. Connect one tubing connector to the other end of the main tubing as shown. There are 2 split tubes with tubing connectors attached on each end. Connect the open end (without the tubing connectors) to the Y-Connector openings as shown. Insert the other sides of the tubing with the tubing connectors into the air vents on each diaphragm caps as shown.

  1.  Make sure the tubing set is assembled correctly
  2. When one side of the split tubes is removed and the plug is inserted into the opening on the Y-Connector, it becomes a single pump.

Crane-EE-9002-Double-Electric-Cordless-Rechargeable-Breast-Pump-User-Manual-Image-7

Connecting the Pump Unit

Insert the main tube with the single tubing connector into the right side of the pump unit. Ensure that the pump unit and the expression collection kit are connected tightly, otherwise suction may be affected. To connect the power supply, Plug the AC Adaptor into the wall outlet. Attach the power cord to the left side of the pump unit.

Crane-EE-9002-Double-Electric-Cordless-Rechargeable-Breast-Pump-User-Manual-Image-8

Cleaning, Hygiene & Care

How to keep your pump clean
Follow the instructions for cleaning, caring and storing for your Crane breast pump and parts. Failure to do so may pose a health risk and void the warranty. Your Crane pump kit is designed for single use only, it is not intended for multiple users. Every mother should have their own pump kit. Use multiple users risks the spread of transmissible diseases. Make sure to wash your hands with soap and warm water prior to touching the breast pump and the breast pump parts. Avoid touching the inner parts of the milk collection kit parts which come in to contact with breast milk. The motor unit must never be exposed or immersed into water or other liquids for cleaning. The motor unit must not be Sterilized in a microwave sterilizer.

Bottle Kit:
Bottles should be sterilized after every use with babies under 3 months of age. Wash and sanitize the Bottle Kit before your first use. Sterilize daily and wash after each use.

Milk Collection Kit:
Wash and sanitize the Milk Collection Kit before your first use. Sanitize daily and wash after each use. The breast pump milk collection kit has a closed system therefore cleaning the tubes is not necessary. If you happen to see moisture in the tubes alter pumping, remove the tubes from the breast shields. Turn on the motor for a few minutes until the moisture is gone.

Pump unit:
Wipe pump unit with clean damp cloth. Sterilize the unit using a disinfectant wipe and wipe off the outside of the tubing and the tubing jack (where tubing attaches to the pump unit). NEVER submerge the pump unit in water or use in a sterilizer.

STERILIZING YOUR PUMP PARTS BEFORE FIRST USE AND ONCE DAILY

  1.  Place disassembled breast shield, silicone cushions, diaphragm, bottle and bottle parts that are safe to boil into a pot and cover with water. DO NOT BOIL TUBES AND CONNECTORS.
  2.  Put the pot over the heat and bring to a boil.
  3.  Boil for 3-5 minutes.
  4.  Allow water to cool and gently remove items with clean tongs
  5. Do not boil tubes or connectors
  6. Allow to air-dry thoroughly. Place sanitized pump parts and bottle brush on a clean, unused dish towel or paper towel in an area protected from dirt and dust. Do not use a dish towel to rub or pat items dry because doing so may transfer germs to the items.
  7.  Store dry parts in a cool, clean place when not in use. DO NOT store wet or damp parts.

Before Each Use
Wash your hands well with soap and water and dry thoroughly before touching any part of your pump.

  •  Make sure that all parts are totally dry prior to use. Wet or damp parts will cause poor suction.
  • Once all parts are totally dry, re-assemble the milk collection kit.
  •  Make sure that you check for wear and tear on tubes, plastic and silicone parts.
  •  Replace worn out or damaged parts immediately as this will affect suction.

CLEANING THE TUBES
The Crane Double Electric Breast Pump is a closed system, therefore cleaning the tubes is not necessary. If you see moisture in the tubes after pumping, remove the tubes from the breast shields. Turn the unit on for a few minutes until the moisture is gone.

Button & Display Guide

How to use the buttons on your pump unit

Crane-EE-9002-Double-Electric-Cordless-Rechargeable-Breast-Pump-User-Manual-Image-9

  1. Power – turn unit in and off when plugged into an outlet
  2. Phase Function (stimulation/expression phase Icons)- press this to change between functions Tap to cycle through the different expression phases. There are 2 phase options for pumping, each phase is represented by these icons on the display when selected.
  3.  +/- increase and decrease suction- press + to increase and – to decrease suction in each function.
  4. Level (Setting) of suction (?) indicator (1-9 levels)
  5.  Battery Power Indicator
  6.  Minutes remaining in 30 minute cycle

Ready to Pump

  1.  On/Off Switch- turn unit to ON/OFF when pumping without unit being plugged into a wall outlet. Ensure unit is turned OFF after use to prolong the battery life.
  2.  Non slip feet to secure unit placement for hands free pumping

Crane-EE-9002-Double-Electric-Cordless-Rechargeable-Breast-Pump-User-Manual-Image-10

Using Your Crane Pump

  • Position the breast shield – Massage around your breast, then position your nipple in the center of the silicone 3D cushion.
  • Pumping Mode – Switch on your pump, starting with the stimulation mode iconto encourage letdown. Once your milk begins to flow (usually after a couple of minutes), switch to an expression mode icon

Finding the right settings for you:

  •  When using your pump for the first few times, experiment with the different expression setting (-/+) to find the one that works well for you.
  •  If your milk flow begins to drop, you may find it helpful to gently massaging your breast.
  •  Adjust the suction level gradually to one that is effective but still comfortable. Don’t choose a level that’s too high and uncomfortable.
  •  You may find it helpful to switch between the Expression icon and stimulation icon Phases while you are pumping.
  • How long to pump: Use your pump to express your milk until the breast is empty, which is usually between 15-30 minutes.
  • Finishing your pumping session: When you are ready to stop expressing, switch your Crane pump off. Ensure the pump motor is OFF before removing the milk collection kit from your breasts.
  • Time to clean: Please see the instructions for rinsing and washing or sterilizing your pump parts that have come in contact with your breast or breast milk. Wipe your pump unit.

Storing & Using Breast Milk

  • Please check the latest advice on storing and using thawed breast milk with your physician or lactation consultant.
  • Always keep your stored breast milk in a sterilized bottle or sterile milk storage bag, adding the date if you are planning to freeze it. Breast milk can be kept at room temperature for 4 hours.
  • You can store your breast milk in the fridge for up to 48 hours at 2-4 degrees C.
  • Choose the coldest part of the fridge, such as above the vegetable drawer, not in the door. Your breast milk can be stored for up to 3-6 months in a freezer. Fill your storage bag or bottle no more than ¾ full, as the frozen milk will expand.
  • Allow your frozen breast milk to defrost itself in the fridge. When it is thawed, use it straight away.
  • Your baby might like warm milk. Stand the bottle in a container of hot water and let it heat to body temperature before feeding. Microwaving bottles is not recommended because the uneven heating can burn your baby’s mouth, and the heat can damage the food value and health protecting qualities of breast milk.
  • Once warmed, and before feeding, always swirl the container of the milk to mix well and test the temperature by shaking a few drops of milk on the inside of your wrist. The milk should feel just warm on your skin

 Troubleshooting Guide

CauseSolution
Low Suction
  •   Air Leakage of connecting parts (valve, cushion, tubing, or connector).
  •    Pump unit’s air circulator is blocked. Tubing is bent or residue left in tubing.
  •  Check each part, make sure connections are on correctly and securely.
  •   Untangle tubing.
  •   Clean or replace tubing if needed.
No Suction
  •     Valve is not assembled in correct place or the valve is deformed and damaged.
  • Ensure valve is securely attached to the breast shield body.
No Power
  •       The AC adaptor Is not plugged into the main socket or the DC socket on the pump unit
  •       The rechargeable battery is flat
  •    Ensure AC Adaptor is plugged in to the wall outlet and the DC socket on the unit.
  • Charge the battery.
Milk or Water in Tubing
  • Diaphragm is not correctly positioned or is damaged
  •      Stop immediately. Clean, reposition, or replace the diaphragm before pumping. Dry or replace tubing before pumping.
Display Shows E1 or E2 Alarm
  •   E1: Open Circuit
  •    E2: Short Circuit
  •   Please contact your Crane Customer Care.

Technical Data

  • AC adapter: Input 100-240V ~ 50/60Hz
  • Output: 10V ~ 1.5A
  • Li-ion polymer battery power: 7.4V ~ 1750mAh
  • Total power: 15W
  • Electronic equipment protection: Type II class
  • Stimulation speed: 70-100T/min
  • Stimulation suction: – 0.012MPa (MAX) (9 steps)
  • Expression speed: 24-70T/min
  • Expression vacuum: -0.026MPa (MAX) (9 steps)
  • Timer: 30 minutes

Leave a Reply