Vibeat-MD300C318T-Bluetooth-Fingertip-Pulse-Oximeter-Feature

Vibeat MD300C318T Bluetooth Fingertip Pulse Oximeter User Manual

Vibeat-logo

Vibeat MD300C318T Bluetooth Fingertip Pulse Oximeter User Manual

Vibeat-MD300C318T-Bluetooth-Fingertip-Pulse-Oximeter-Fig8

Safety Information

Please read this manual thoroughly before using the pulse oximeter! Keep it in hand for furture reference. Warnings alert the user to potential serious outcomes, such as injury or adverse events to the patient or user. Cautions alert the user to exercise care necessary for the safe and effective use of the pulse oximeter. Notes contain important information that may be overlooked or missed.

Warnings!

  • DO NOT strike or needle the battery.
  • Keep away from source of fire and/or heat.
  • DO NOT disassemble the oximeter or its accessories.DO NOT use the pulse oximeter in an MRI or a CT environment.
  • DO NOT use the pulse oximeter in the presence of flammable anesthetics.
  • Explosion hazard: DO NOT use the pulse oximeter in an explosive atmosphere.
  • Chemicals from a broken OLED panel are toxic when ingested. Use caution when the oximeter has a broken display screen.
  • The pulse oximeter is intended only as an adjunct in patient assessment. It must be used in conjunction with other methods of assessing clinical signs and symptoms.
  • Check the pulse oximeter application site frequently to determine the positioning of the measurement and circulation and skin sensitivity of the patient.
  • Prolonged use or the patient’s condition may require changing the measurement site periodically. Change measurement site and check skin integrity, circulatory status, and correct alignment at least every 4 hours.
  • Follow local ordinances and recycling instructions regarding disposal or recycling of the device and device components, including batteries.
  • Only the authorized service personnel can replace the battery or repair this device. This device uses a fixed lithium-ion battery inside. Do not try to replace the battery by yourself at any time. For longer battery life, only charge the battery when the battery power is empty.
  • This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.

Cautions!

  • Inaccurate measurements may be caused by autoclaving, ethylene oxide sterilizing, or immersing the sensors in liquid may cause inaccurate readings.
  • Significant levels of dysfunctional hemoglobin (such as carbonyl-hemoglobin or methemoglobin) may cause inaccurate readings.
  • Intravascular dyes such as indocyanine green or methylene blue may cause inaccurate readings.
  • SpO2 measurements may be adversely affected in the presence of upper ambient light. Shield the sensor area (with a surgical tower, or direct sunlight, for example) if necessary.
  • Excessive patient movement may cause inaccurate readings.
  • Upper-frequency electrosurgical interference may cause inaccurate readings.
  • Venous pulsations may cause inaccurate readings.
  • Placement of a sensor on an extremity with a blood pressure cuff, arterial catheter, or intravascular line may cause inaccurate readings.
  • The patient has hypotension, severe vasoconstriction, severe anemia, or hypothermia may cause inaccurate readings.
  • Operation of the pulse oximeter may be affected by the use of an electrosurgical unit (ESU).
  • The pulse oximeter must be able to measure the pulse properly to obtain an accurate SpO2 measurement. Verify that nothing is hindering the pulse measurement before replying on the SpO2 measurement.
  • The patient is in cardiac arrest or is in shock may cause inaccurate readings.
  • Fingernail polish or false fingernails may cause inaccurate SpO2 readings.
  • Federal Law (U.S.A) restricts this device to sale by or on the order of a physician.
  • All CHOICEMMED devices are designed to be comliant with rules and regulations in locations they are sold and will be labelled as required.
  • Any changes or modifications to CHOICEMMED equipment, not expressly approved by CHOICEMMED, could void the user’s authority to operate the equipment.

CHAPTER 1 General Information

Oxygen saturation is a percentage of oxyhemoglobin (HbO2) capacity, compounded with oxygen, by all combinative hemoglobin (Hb) capacity in blood. In other words, it is consistency of oxyhemoglobin in blood. It is a very important parameter for the Respiratory Circulation System. Many respiratory diseases can result in low oxygen saturation in human blood. Additionally, the following factors can reduce oxygen saturation: Automatic regulation of organ dysfunction caused by Anesthesia, Intensive Postoperative Trauma, injuries caused by some medical examinations. That situation might result in light-headedness, asthenia, and vomiting. Therefore, it is very important to know the oxygen saturation of a patient so that doctors can find problems in a timely manner. The pulse oximeter integrated with Bluetooth® wireless technology is designed to monitor vital signs, such as SpO2 and pulse rate anytime, anywhere.

Measuring Principle

An experience formula of data process is established taking use of Lambert Beer Law according to Spectrum Absorption Characteristics of Reductive hemoglobin(HbR) and Oxyhemoglobin (HbO2) in red light and near-infrared light zones. The photoelectric oxyhemoglobin inspection technology is adopted in accordance with capacity pulse scanning and recording technology, so that two beams of different wavelength of lights (660nm red light and 940nm near infrared light) can be focused onto human nail tip through perspective clamp finger type sensor. Then measured signal can be obtained by a photosensitive element, information acquired through which will be shown on the display through process in electronic circuits and microprocessor.

Diagram of Operation Principle:

Vibeat-MD300C318T-Bluetooth-Fingertip-Pulse-Oximeter-Fig1

  1. Red and Infrared light emission diodes
  2. Red and Infrared photodiode

Product Features

  • Color OLED screen displays SpO2, PR, waveform and pulse bar; 4 display modes.
  • Rechargeable lithium-ion battery, battery charging via USB cable or adapter.
  • Real-time and non real-time Bluetooth® data transmission.
  • Real-time inner clock.
  • Software upgrade with Bootloader.
  • Suitable for pediatrics and adults.

Intended Use

The MD300C318T Fingertip Pulse Oximeter is intended for continuous use or spot checking in measuring and displaying functional arterial oxygen saturation (SpO2) and pulse rate of patients in hospitals and home care. It is intended for adult and pediatric patients on finger between 0.3-1.0 inch (0.8 – 2.5 cm) thick.

Appearance Introduction 

Vibeat-MD300C318T-Bluetooth-Fingertip-Pulse-Oximeter-Fig2

Description of the Device

  • Display screen: provides an organic light-emitting diode (OLED) display panel. [Figure 1.2]
  • Power switch & function button is used to turn on the device and set the display orientation [Figure 1.2-2]. See full description of the functions below, under items 1-10.
  • USB cable port: is used to charge the device and transfer data. [Figure 1.3] The display screen is illustrated in the figure below

Vibeat-MD300C318T-Bluetooth-Fingertip-Pulse-Oximeter-Fig3

Description of fig1.4

  1. Time display: The time shown in the figure is 11:45.
  2. SpO2% refers to the oxygen saturation in blood.
  3. Tabulation of the measured SpO2% reading. Here, it is valued at 98%.
  4. This field indicates SpO2% Plethysmograph signal’s dimensions.
  5. This field indicates the measurement of pulse rate (PR). Here, the PR is valued at 60 beats per minute (bpm).
  6. The pulse graph bar indicates pulse amplitude.
  7. PRbpm refers to Pulse Rate beats per minute.
  8. This field indicates the level of battery power in the pulse oximeter. When battery power is low, the icon will be empty and turn to red.
  9. Bluetooth® transmission indicator.
  10. Brightness level indicator.

NOTE: The illustrations used in this manual may differ slightly from the appearance of the actual product.

Description of Symbols

Vibeat-MD300C318T-Bluetooth-Fingertip-Pulse-Oximeter-Fig4

CHAPTER 2 Settings

Configuring Your Pulse Oximeter

There are some settings which can be adjusted by software, including date and time, switching, real-time and non realtime data transmission modes, powering off, and updating the software of the pulse oximeter.

Selecting a Display Mode

You can adjust the display mode of your oximeter when performing SpO2 and pulse rate measurements, every time you press the button, the pulse oximeter will switch to another display mode. During the measuring, each time you press the button the oximeter will switch to another display mode, there are 4 display modes shown as below:

Vibeat-MD300C318T-Bluetooth-Fingertip-Pulse-Oximeter-Fig5

CHAPTER 3 Take a Measurement

  • Open the clamp and insert a finger into the oximeter, as illustrated in the right figure. Then gently release the clamp.
  • Make sure that the distance between you and other Bluetooth® device within 10m. (30 ft.)

Vibeat-MD300C318T-Bluetooth-Fingertip-Pulse-Oximeter-Fig6

  • Press the power button, the display will light up and after a few seconds, the device will display SpO2% and pulse rate measuring values. The oximeter will also transmit the recorded information via Bluetooth® to other Bluetooth® device.
  • After transmitting data to other Bluetooth® device, the oximeter will shut down automatically.

NOTES:

  • Keep your tested hand still during the measurement.
  • Do not pour or spray liquids onto the oximeter, and do not allow any liquid to enter any openings in the device. Make the oximeter dry thoroughly before reusing.
  • The oximeter will shut down after a period of time if no measurement is being taken.
  • Clean the rubber touching the finger probe inside of the pulse oximeter with a soft cloth dampened with 70% isopropyl alcohol, and cleans the test finger probe using alcohol before and after each measurement.

CHAPTER 4 Battery Charge

Power Supply

  • Battery Model: SP080
  • Battery Type: One 3.7V Lithium Ion Rechargeable Battery

Battery Charge

  • Connect the oximeter with the attached USB cable, shown as the following figure.
  • Connect the other end of the USB cable to the attached charger and plug the charger into an electrical outlet.
  • Charging will complete in 4-5 hours. The battery indicator will stop flickering if the charging process is completed.

Vibeat-MD300C318T-Bluetooth-Fingertip-Pulse-Oximeter-Fig7

  • Charging temperature: 0℃~40℃
  • Charging Voltage: 110Vac~240Vac

Warning!

  • Avoid strongly impacting to the oximeter.
  • Avoid the oximeter to be exposed straightly to sunlight.
  • Only use the attached battery charger. Otherwise, damage even danger may be caused to oximeter or person.
  • Do not charge the battery for more than 5 hours, or that may cause damage to the battery.
  • Do not charge or preserve the battery in too hot or cold environment, the proper temperature is 0~40℃ for charging
  • Please refer to details in Chapter 6 Specifications.
  • Charging is not recommended when using the oximeter

CHAPTER 5 Data Transmission

The measurement results saved in the pulse oximeter can be uploaded to other Bluetooth® devices for review or management. The modes of data transmission include Realtime Bluetooth® transmission and Non-realtime Bluetooth® transmission. Please refer to “MD300C318T Communication Protocol” for details of data transmission.

Realtime Bluetooth® transmission(default):

  • Bluetooth® icon will flash when MD300C318T is not connected.
  • Bluetooth® icon will not flash when the valid pairing of MD300C318T is connected.
  • If Bluetooth® is not connected, the measured data will not be saved and MD300C318T will be in a wait state .(If Bluetooth® has not been connected for 3 minutes, MD300C318T will shutdown automatically.)
  • If Bluetooth® connection is successful, MD300C318T will upload the values of blood oxygen saturation, pulse rate, blood oxygen volume wave, battery power to other Bluetooth® devices; Meanwhile, MD300C318T can respond and execute commands from other Bluetooth® devices. (Please see MD300C318T Communication Protocol.)

Non-realtime Bluetooth® transmission:

The data transimission screen is shown in Fig.5.1.

Vibeat-MD300C318T-Bluetooth-Fingertip-Pulse-Oximeter-Fig8

  • MD300C318T will automatically judge the effective measured data ( 6 single point effective measurements will be stored)
  • Bluetooth® will automatically open and upload the measured data after measurement.
  • MD300C318T software system can set the number of stored data of one successful measurement, the number ranging from 1-10.
  • If data upload fails, the measured data will be saved automatically and uploaded with next measured data. At most 100 pieces of failed uploaded data can be stored and re-uploaded.
  • The distance between the pulse oximeter and other Bluetooth® device should be not more than 10m.
  • The measurement data is saved every 4s and at most 100 pieces of data records can be saved in the pulse oximeter.

CAUTION: The Bluetooth® transmission may be influenced in different environments, such as walls, metallic doors, steel wire netting and an MRI or a CT environment and so on.

CHAPTER 6 Specifications

Display:
Display type: OLED display
SpO2 display range: 0~99%
SpO2 measuring range: 70~99%
PR display range: 0~235 BPM
PR measuring range: 30~235 BPM
PR display mode: Amplitude Bar
Data update period: <15 s
Resolution:
SpO2%: 1%
Pulse rate: 1BPM
Measurement Accuracy:
SpO2: 70%~99% ±3%; ≤69% unspecified.
PR: 30~99 bpm ±2 bpm; 100~235 bpm ±2%.
Antenna Information:
Antenna Type/Patten: Internal
Frequency Range: 2402 to 2480 MHz
LED Wavelengths and Output Power:
Red: approximately 660nm @0.8mW maximum average
Infrared: approximately 940nm @0.8mW maximum average
Bluetooth®
The device name format of MD300C318T is “B *********”.
(“*********” denotes numbers from 0 to 9, for example
B110500001.) The device name will be labeled on the
back of MD300C318T pulse oximeter.
Wireless transmission distance: 0~10m
Bluetooth® default paired password: 0000.
Physical Characteristics:
Dimensions: 60mm x 36mm x 38mm (2.36”x 1.42”x
1.50”)
Weight: 50g (including the lithium ion battery)
Battery Life:
Typical lithium ion battery: 300 cycles
Charge temperature: 0℃~40℃
Environment Conditions:
Operating Temperature: 5℃~40℃
Storage Temperature: -20℃~-55℃
Humidity: ≤80%, no condensation in operation
≤93%, no condensation in storage

Accessories:

  • AC/DC Adapter·····························································································································································1piece
  • Instruction Manual ························································································································································1piece
  • Lanyard·········································································································································································1piece
  • USB cable·············1piece

CHAPTER 7 Maintenance and Calibration

  • Please charge the battery in time when the low power indicator flickering.
  • Clean the surface of the fingertip oximeter before it is used in diagnosis for patients.
  • It is better to preserve the product in the environment specified in this manual.
  • It is recommended that the product should be kept in a dry environment. A wet ambient might affect the lifetime, even damage the device.
  • Please follow the law of local government to deal with the used oximeter and the accessories.

CHAPTER 8 Troubleshooting

Problems Reasons Solutions
SpO2% or pulse rate does not display.

  • Finger is not plugged correctly.
  • Patient’s SpO2 value is too low to be measured.
  • Retry by plugging the finger.
  • There is excessive illumination.
  • Measure other patients to make sure that no problem exists in the product. Go to a hospital in a timely manner for an exact diagnosis.
  • SpO2% or PR is shown unstably.
  • Finger might not be plugged deep enough into the clamp probe.
  • Excessive patient movement.
  • Retry by inserting the finger to the end.
  • Stop moving the finger, hand or body.
  • The oximeter cannot be powered on.
  • No battery or low power of battery.
  • Batterymight be installed incorrectly.
  • The Monitor might be damaged.
  • Please replace battery.
  • Please reinstall the battery.
  • Please contact with local customer service centre.
  • Display suddenly turns off.
  • The oximeter is automatically powered off when no signal is detected longer than 8 seconds.
  • The batteries power is too low to work.
  • Relocate the probe on another finger or restart the oximeter and be sure the signal strength is strong enough for stable display.
  • Replace the battery.

Error 1
Error 2
Error10
ROM error
RAM error
EEPROM damaged or dry joint
Please contact with local customer service centre.
Please contact with local customer service centre.
Please contact with local customer service centre.

REFRENCE LINK

https://vibeatstore.com/

Leave a Reply